Sample Sheet Generation Issue and CLPA Issues When Samples Have Been Assigned QC Flag Prior to Entering Steps
Last updated
Last updated
Last Updated: July 2024
The following issues will happen if any of the samples in the pools has been assigned QC flag in prior steps when using impacted integrations (see Impact to Customer):
Sample sheet and planned run generation will fail
Clarity LIMS Product Analytics (CLPA) payload generation (pooling step only) will fail
CLPA association of libraries with sequencing run will fail
Impacted Integration Packages
NextSeq1k2k INT v2.4.0 and v2.5.0
NovaSeq X INT v1.1.0 and v1.2.0
UPA Script 1.2.0, 1.2.1 and 1.3.0
Scenarios
NextSeq1k2k INT v2.5.1
NovaSeqX INT v1.2.1
UPA INT v1.3.1
Integration(s)
Scenario
Impact
Recommendations
NextSeq1k2k INT v2.4.0 and v2.5.0
NovaSeq X INT v1.1.0 and v1.2.0
One or more samples in the pool(s) have been assigned with QC flag before entering these steps:
Step with automation to generate sample sheet and planned run
Unable to generate sample sheet and planned run
NS1k2k INT:
Downgrade to NS1k2k INT v2.3.1
NovaSeq X INT:
No workarounds
UPA Script 1.2.0, 1.2.1 and 1.3.0
One or more samples in the pool(s) have been assigned with QC flag before entering these steps:
Pooling steps
Sequencing run steps
CLPA fails to generate pooling step payload
CLPA fails to associate library with sequencing run
No workarounds
Recommend to update all affected integration packages to the latest version once available
Recommend to update all affected integration packages to the latest version once available