The integration includes the following:
Automated beadchip accessioning in Clarity LIMS (via Illumina Connected Analytics during pre-imaging verification).
Automated verification of beadchip status. Samples are automatically created upon successful verification.
Automated beadchip imaging status tracking (imaging run started; errored out; completed successfully; or ended by user events).
Manual invocation of analysis on Illumina Connected Analytics.
Automated analysis status (for completed successfully or errored out events) tracking.
The Illumina iScan Integration Package v1.1.0 supports the integration of Clarity LIMS to iScan systems. The package includes the following workflows, which are included in the Illumina Preset Protocols (IPP) Package v2.6.0:
iLASS Infinium Genotyping Assay v1.0
iLASS Infinium Genotyping Assay with PGx v1.0
This integration enables you to do the following tasks:
Accession the BeadChip.
Handle the imaging run event (including pre-imaging verification for the imaging step).
Handle the analysis run event for the analysis step.
The configuration used in this integration supports iScan lab processes. Any changes to the configuration (including renaming protocols, steps, and fields) can cause issues with these processes.
In the iScan workflows, BeadChips are accessioned into Clarity LIMS before the hybridization step of the post-lab protocol. The BeadChip container type is defined based on the BeadChip product type (eg, PGx) and the product version. Because every BeadChip has a unique barcode, Clarity LIMS can track usage throughout the entire workflow. The following BeadChip names are supported in iScan Integration v1.1.0:
GCRA-24v1-0
GDA_PGx-8v1-0
GSA-48v4-0
If you need help accessioning your BeadChip, contact Illumina Support.
Master Step: Image Infinium BeadChip on iScan (iLASS Infinium Genotyping Assay v1.0) or Image Infinium BeadChip on iScan (iLASS Infinium Genotyping Assay with PGx v1.0)
The following table lists field configuration details defined on the Image Infinium BeadChip on iScan (iLASS Infinium Genotype Assay/with PGx v1.0) master step. These fields are located under Configuration, on the Master Step Fields tab.
When BeadChips are loaded onto the iScan system, the iScan Control Software (iCS) with LIMS mode enabled performs pre-imaging verification to make sure that user loads the correct BeadChip. This verification is done as follows.
iScan Control Software verifies the BeadChip container is queued to the imaging step in Clarity LIMS.
After the verification completes sucessfully, Clarity LIMS iScan integration creates run on ICA and starts BeadChip imaging starts automatically.
When the BeadChip is scanned with an iScan, Clarity LIMS receives the events of the run status update. Then, the custom fields are automatically populated with information.
The Run Status field can show the following statuses:
Running
Completed
Aborted
Failed
If there is a run error event, the Run Complete Time field is not populated. The Failure Reason field will be populated only when the run error event happens.
Master Step: Invoke Infinium Analysis (iLASS Infinium Genotyping Assay v1.0) or Invoke Infinium Analysis (iLASS Infinium Genotyping Assay with PGx v1.0)
In this step, the iScan Integration handles events related to BeadChip analysis. This step is a manual step that requires data input of Analysis Configuration ID and Analysis Run Name fields to start the analysis run.
The following table lists field configuration details defined on the Invoke Infinium Analysis (iLASS Infinium Genotyping Assay/with PGx v1.0) step. These fields are located under Configuration, on the Master Step Fields tab.
On the Record Details milestone, navigate to Step Details and enter the Analysis Run Name. The Analysis Run Name can only contain a maximum of 255 alphanumeric characters, underscores, hyphens, and periods. The name cannot start with the wfr. characters.
Enter the Analysis Configuration ID. To retrieve the configuration ID that is associated with the analysis, contact Illumina Support.
Select Start Analysis v1.0 to start the analysis run.
When the analysis is started on ICA, Clarity LIMS receives the event and displays related information. The custom fields are automatically populated with information.
The Analysis Status field can show the following statuses:
Analysis Running
Analysis Completed
Analysis Completed with Warning
Analysis Aborted
Analysis Timedout
Analysis Errored
If there is an analysis error event, the Analysis Complete Time field is not populated.
This step is automated and must not be started manually.
Do not delete, rename, or modify. These actions will break the integration.
The Illumina iScan integration provides preconfigured workflows that map to established lab protocols and steps used with the iScan system. The integration also supports the associated consumables and assays used in these protocols.
Field Name
Field Type
Options
Additional Options and Dropdown Items
BeadChip Barcode
Text
Read Only
Instrument
Text
Read Only
Run Start Time
Text
Read Only
Data Path
Text
Read Only
Instrument Run ID
Text
Read Only
Run Status
Text
Read Only
DMAP Location
Text
Read Only
Instrument Type
Text
Read Only
Failure Reason
Text
Read Only
Log
Multiline Text
Read Only
ICA Run ID
Text
Read Only
Run Complete Time
Text
Read Only
Field Name
Field Type
Options
Additional Options and Dropdown Items
Analysis Complete Time
Text
Read Only
Analysis Run Name
Text
Required
Analysis Configuration ID
Text
Required
Analysis Start Time
Text
Read Only
Analysis Report
Text
Read Only
Analysis Status
Text
Read Only
Analysis Run ID
Text
Read Only
Log
Multiline Text
Read Only
Last Updated: November 2024
Release Date: July 2024
Document Version: 2
These release notes describe the key changes to software components for the Clarity LIMS iScan Integration Package version 1.2.0.
Refer to under Instruments & Integrations.
Updated iScan integration to support 'ILASS Infinium Genotyping Assay v1.1' and 'ILASS Infinium Genotyping Assay with PGx v1.1' workflows
Supports the new 'Infinium Methylation Array EX-48' beadchip
Removed unused Analysis Result column from Sample Details table in Infinium Analysis step.
Integration-related properties can now be accessed and updated via System Setting in Clarity v6.3. Refer to for configurable properties.
Updated integration properties to be extensible to future array projects
Replaced supportedImagingWorkflowStepNames with supportedImagingWorkflowStepNamePattern
Replaced supportedAnalysisWorkflowStepNames with supportedAnalysisWorkflowStepNamePattern
Added analysis run ID in analysis run logging to improve troubleshooting.
Fixed imaging result not reflecting correctly in the "Infinium Array Imaging on iScan" step upon re-imaging of beadchip.
Updated iScan install details to table retrieved by getInstallHistory script.
Warning log message displays when the ImagingAborted and ImagingFailed events are processed.
The Illumina iScan Integration Package v1.2.0 supports the integration of Clarity LIMS to iScan systems. The package includes the following workflows, which are included in the Illumina Preset Protocols (IPP) Package v2.9.0:
iLASS Infinium Genotyping Assay v1.1
iLASS Infinium Genotyping Assay with PGx v1.1
This integration enables you to do the following tasks:
Accession the BeadChip.
Handle the imaging run event (including pre-imaging verification for the imaging step).
Handle the analysis run event for the analysis step.
The configuration used in this integration supports iScan lab processes. Any changes to the configuration (including renaming protocols, steps, and fields) can cause issues with these processes.
Beadchip are auto accessioned for the workflows listed in Supported Workflows. The beadchip containers are created in Clarity by iLASS at the hybridization step of the post-lab protocol. Prior to imaging, a verification is performed and beadchips that are not accessioned prior to start of the workflows will be automatically accessioned by Illumina Connected Analytics (ICA). The BeadChip container type is defined based on the BeadChip product type (eg, PGx) and the product version. Because every BeadChip has a unique barcode, Clarity LIMS can track usage throughout the entire workflow. The following BeadChip names are supported in iScan Integration v1.2.0:
GCRA-24v1-0
GDA_PGx-8v1-0
GSA-48v4-0
MSA-48v1-0
Master Step: Image Infinium BeadChip on iScan (iLASS Infinium Genotyping Assay v1.1) or Image Infinium BeadChip on iScan (iLASS Infinium Genotyping Assay with PGx v1.1)
The following table lists field configuration details defined on the Image Infinium BeadChip on iScan master step. These fields are located under Configuration, on the Master Step Fields tab.
When BeadChips are loaded onto the iScan system, the iScan Control Software (iCS) with LIMS mode enabled performs pre-imaging verification to make sure that user loads the correct BeadChip. This verification is done as follows.
iScan Control Software verifies the BeadChip container is queued to the imaging step in Clarity LIMS.
After the verification completes sucessfully, Clarity LIMS iScan integration creates run on ICA and starts BeadChip imaging starts automatically.
When the BeadChip is scanned with an iScan, Clarity LIMS receives the events of the run status update. Then, the custom fields are automatically populated with information.
The Run Status field can show the following statuses:
Running
Completed
Aborted
Failed
If there is a run error event, the Run Complete Time field is not populated.
Master Step: Invoke Infinium Analysis (iLASS Infinium Genotyping Assay v1.1) or Invoke Infinium Analysis (iLASS Infinium Genotyping Assay with PGx v1.1)
In this step, the iScan Integration handles events related to BeadChip analysis. This step is a manual step that requires data input of Analysis Configuration ID and Analysis Run Name fields to start the analysis run.
The following table lists field configuration details defined on the Invoke Infinium Analysis master step. These fields are located under Configuration, on the Master Step Fields tab.
On the Record Details milestone, navigate to Step Details and enter the Analysis Run Name. The Analysis Run Name can only contain a maximum of 255 alphanumeric characters, underscores, hyphens, and periods. The name cannot start with the wfr. characters.
Enter the Analysis Configuration ID. To retrieve the configuration ID that is associated with the analysis, contact Illumina Support.
Select Start Analysis v1.0 to start the analysis run.
When the analysis is started on ICA, Clarity LIMS receives the event and displays related information. The custom fields are automatically populated with information.
The Analysis Status field can show the following statuses:
Analysis Running
Analysis Completed
Analysis Completed with Warning
Analysis Aborted
Analysis Timedout
Analysis Errored
If there is an analysis error event, the Analysis Complete Time field is not populated.
Last Updated: November 2024
Release Date: September 2023
Document Version: 2
These release notes describe the key changes to software components for the Clarity LIMS iScan Integration Package version 1.1.0.
Refer to under Instruments & Integrations.
Updates to the Java and third-party dependency libraries used in the integration.
Fixed the SIS core service not being properly registered in systemctl.
Warning log message displays when the ImagingAborted and ImagingFailed events are processed.
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging is started, completed, aborted, or has failed
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging fails
Automatically populated when imaging is started, completed, aborted, or has failed
Automatically populated when imaging is started
Automatically populated when imaging is completed or aborted
Automatically populated when analysis is completed or aborted
Automatically populated when analysis is started
Automatically populated when analysis is completed
Automatically populated when analysis is started, completed, aborted, or has failed
Automatically populated when analysis is started
Automatically populated when analysis is started, completed, aborted, or has failed
This step is automated and must not be started manually.
Do not delete, rename, or modify. These actions will break the integration.
Refer to for the properties installed with Illumina iScan Integration Package v1.2.0.
Version | Changes |
2 |
|
1 |
|
Version | Changes |
2 |
|
1 |
|
The integration includes the following:
Automated beadchip accessioning in Clarity LIMS upon beadchip creation on Illumina Connected Analytics (ICA).
Automated verification of beadchip status. Samples are automatically created upon successful verification.
Automated beadchip imaging status tracking (imaging run started; errored out; completed successfully; or ended by user events).
Manual invocation of analysis on ICA.
Automated analysis status (for completed successfully or errored out events) tracking.
The integration utilizes the following pre-configured workflows in IPP:
iLASS Infinium Genotyping Assay v1.0
Field Name | Field Type | Options | Additional Options and Dropdown Items |
BeadChip Barcode | Text | Read Only |
Instrument | Text | Read Only |
Run Start Time | Text | Read Only |
Data Path | Text | Read Only |
Instrument Run ID | Text | Read Only |
Run Status | Text | Read Only |
DMAP Location | Text | Read Only |
Instrument Type | Text | Read Only |
Failure Reason | Text | Read Only |
Log | Multiline Text | Read Only |
ICA Run ID | Text | Read Only |
Run Complete Time | Text | Read Only |
Comment | Multiline Text |
Field Name | Field Type | Options | Additional Options and Dropdown Items |
Analysis Complete Time | Text | Read Only |
Analysis Run Name | Text | Required |
Analysis Configuration ID | Text | Required |
Analysis Start Time | Text | Read Only |
Analysis Report | Text | Read Only |
Analysis Status | Text | Read Only |
Analysis Run ID | Text | Read Only |
Log | Multiline Text | Read Only |
The iScan workflow includes the following functionality:
Preset protocol.
Ability to attach the scan log to the samples.
Protocol Type = Other
Next Steps Configuration
Master Step Name = Image Infinium BeadChip - iScan v1.0
Step Type = No Outputs
Defaults
Sample Grouping = Group by Containers
Well Sort Order = Row
Sample Table (Column Headers)
Step Data (Master Step Fields)
Step File Placeholders
Log - Automatically attached
Scan Log - Manually uploaded
Sample Table
Sample Display Default = Collapse
Well Sort Order = Row
Table Columns - Global Fields
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging is started, completed, aborted, or has failed
Automatically populated when imaging is started
Automatically populated when imaging is started
Automatically populated when imaging fails
Automatically populated when imaging is started, completed, aborted, or has failed
Automatically populated when imaging is started
Automatically populated when imaging is completed or aborted
Automatically populated when analysis is completed or aborted
Automatically populated when analysis is started
Automatically populated when analysis is completed
Automatically populated when analysis is started, completed, aborted, or has failed
Automatically populated when analysis is started
Automatically populated when analysis is started, completed, aborted, or has failed
The workflow version and step version for the routing script may vary depending on the version of the IPP.
Category
Field Name
Field Type
Options
Additional Options and Dropdown Items
Container
Container Name
Built-in
Container
LIMS ID (Container)
Built-in
Container
Well
Built-in
Derived Sample
Sample Name
Built-in
Derived Sample
Waiting
Built-in
Project
Project Name
Built-in
Field Name
Field Type
Options
Additional Options and Dropdown Items
Comment
Multiline Text
Category
Field Name
Field Type
Options
Additional Options and Dropdown Items
Derived Sample
Array Secondary Analysis
Text Dropdown
Presets
GS-GENOTYPING
GS-METHYLATION
Default = GS-GENOTYPING
Derived Sample
BeadChip Barcode
Text
Read Only
Derived Sample
BeadChip Position
Text
Read Only
Derived Sample
DNA Plate Barcode
Text
Read Only
Derived Sample
Infinium Kit Name
Text
Read Only
Derived Sample
Sample Name
Built-in
Derived Sample
Sample_Well
Text
Project
Project Name
Built-in