Last Updated: December 2024
Release Date: December 2024
These release notes describe the key changes to Illumina Preset Protocols (IPP) v2.10.
Refer to Compatibility under Instruments & Integrations.
Added a new MiSeq i100 Series Sequencing v1.0 protocol that supports the new MiSeq i100 Series Integration Package v1.0.
Added a new MiSeq i100 Series On-Prem Sequencing v1.0 protocol that supports the new MiSeq i100 Series On-Prem Integration Package v1.0.
Added a new NovaSeq X Series On-Prem Sequencing v1.0 protocol that supports the new NovaSeq X Series On-Prem Integration Package v1.0.
The updated NovaSeq X Series Sequencing v1.2 protocol supports NovaSeq X Series Integration Package v1.3.
The updated NextSeq 500/550 Sequencing v1.3 protocol supports NextSeq 500/550 Integration Package v2.5.
The updated Library Prep Validation v2.3.5 protocol now displays an expanded sample table under the Record Details milestone by default.
Conflicts are expected when upgrading from IPP v2.9 to IPP v2.10 due to new instruments being supported, updates to the configuration settings of BP Aliquot Volume (ul) global field, and bug fixes for NovaSeq X Series Sequencing v1.1 protocol. Refer to IPP Integration v2.10 Manual Upgrade for details.
Importing workflows from IPP on Clarity v6.2 and v6.3 requires longer time (~8 times longer) compared to earlier versions (Clarity v6.1 and below). This issue is fixed in Clarity v6.3.1.
Refer to these additional release notes for new features, updates and defects fixed for instrument integration protocols:
This section describes the manual upgrade process from IPP v2.9 to IPP v2.10 to enable the changes described in Illumina Preset Protocols v2.10 Release Notes. If you do not want to upgrade the latest workflow through IPP, then update the workflow configuration manually using the following information.
The manual upgrade instructions include steps for:
Updating the sequencing instrument presets and routing scripts to support the new MiSeq i100 Series cloud, MiSeq i100 Series on-premise and NovaSeq X Series on-premise integrations.
Updating the precision of BP Aliquot Volume (ul) derived sample global field.
ℹ BP Aliquot Volume (ul) field is used in Illumina Genomics Architecture - NovaSeq Sequencing v2.1, NovaSeq 6000 v3.8, NovaSeqDx v1.2 and NovaSeq X Series Sequencing v1.1 protocols in IPP v2.9.
Updating the Library Prep Validation v2.3.4 protocol:
to display an expanded sample table under the Record Details milestone by default.
to route libraries to new instruments. Refer to these steps for updating the routing scripts.
Updating the NovaSeq X Series Sequencing v1.1 protocol in IPP v2.9 to fix bugs. The steps involve updating container definition, step configuration, automations and global step fields. For manual upgrade instructions for the NovaSeq X Series Sequencing workflow, refer to NovaSeq X Series Integration v1.3.0 Manual Upgrade.
Updating the NextSeq 500/550 Sequencing v1.2 protocol in IPP v2.9 to support generation of forward-orientation samplesheet that is compatible with NextSeq 500/550 Control Software Version 4.2 and Local Run Manager (LRM) v4.0 for selected applications. For manual upgrade instructions for the NextSeq 500/550 Sequencing workflow, refer to NextSeq 500/550 Integration v2.5.0 Manual Upgrade.
In IPP v2.10, new workflows are available. New options for the MiSeq i100 Series (cloud), MiSeq i100 Series (on-premise) and NovaSeq X Series (on-premise) integrations have been added.
This can cause a conflict with the Sequencing Instrument derived sample custom field. The following workflows can be affected:
Illumina DNA Prep (M) Tagmentation v1.0.9
Nextera Mate Pair v1.0.10
Nextera XT DNA v2.0.10
TruSeq DNA PCR-Free v2.0.10
TruSeq Nano DNA v1.0.10
Library Prep Validation v2.3.4
TruSeq ChIP-Seq v1.0.10
TruSeq Methyl Capture EPIC v2.0.11
TruSeq Small RNA v1.0.10
TruSeq Stranded mRNA v2.1.10
TruSeq Custom Amplicon (TSCA) v1.0.10
Nextera Rapid Capture Custom Enrichment v2.0.11
TruSeq DNA Exome v2.0.10
TruSeq RNA Access v2.0.10
TruSight Tumor 170 v2.0.10
When the expected conflicts are encountered during the IPP workflow installation, enter f to import the workflow when prompted. An example of the prompt is as follows.
After the workflow is installed, resolve the conflicts as follows.
From Configuration, select the Custom Fields tab, and then select the Global Fields tab.
Search for Sequencing Instrument.
For Additional Options, add the following to the Dropdown Items list:
NovaSeq X Series On-Prem
MiSeq i100 Series
MiSeq i100 Series On-Prem
Existing library preparation workflows may include routing scripts to integrations. These routing scripts have been updated to add new options for the MiSeq i100 Series (cloud), MiSeq i100 Series (on-premise) and NovaSeq X Series (on-premise) integrations. Modify the automation script as follows.
[Optional] If you are using the MiSeq i100 Series (cloud) instrument, add the instrument to the script:
[Optional] If you are using the MiSeq i100 Series (on-premise) instrument, add the instrument to the script:
[Optional] If you are using the NovaSeq X Series (on-premise) instrument, add the instrument to the script:
[Optional] If you are using the NovaSeq X Series Sequencing workflow v1.2 for cloud, update the workflow version in the script:
[Optional] If you are using the NextSeq 500/550 Sequencing workflow v1.3, update the workflow version in the script:
Select the Configuration tab, and then select the Custom Fields tab.
Under Global Fields, navigate to the Derived Sample section and update BP Aliquot Volume (ul):
Decimal Places Displayed from 0 to 1
Select the Configuration tab, and then select the Lab Work tab.
Select Library Prep Validation workflow.
Select Library Prep Validation protocol.
Select Library Prep Validation step.
Select Record Details milestone.
Toggle Sample Display Default setting to Expand.
The Illumina Preset Protocols (IPP) v2.10 support the integration of Clarity LIMS with established lab protocols.
The integration provides preconfigured workflows that map to lab protocols and steps, and support Illumina library prep kits, reagent kits, assays, and instruments.
This guide provides instructions and describes the components installed in the default configuration of IPP.
In IPP v2.10, new workflows are available. This can cause a conflict with the Sequencing Instrument derived sample custom field for some workflows. Refer to IPP v2.10 Manual Upgrade to resolve conflicts encountered when upgrading from IPP Integration v2.9
IPP Integration v2.10 is distributed as the BaseSpaceLIMS-Illumina-Preset-Protocols RPM package. The RPM package installs the following components:
The illumina-preset-protocols-installer.sh script for listing and importing available workflow configuration package files.
Templates for the following uses:
Sample sheet generation
Sample placement pattern
Covid testing setup and reporting
Illumina Genomics Architecture (IGA)
Automation scripts, including scripts to support IGA and Clarity LIMS Product Analytics (CLPA) automation.
Workflow configuration package files.
IPP Integration v2.10 provides protocols that contain workflow configuration package files. These workflow configuration package files are used to install other workflows and protocols.
QC_Protocols configuration component contains the qc-protocols base configuration. This configuration does not contain any workflows, but it does provide the following protocols that can be incorporated into other workflows:
DNA Initial QC
Library Validation QC
RNA Initial QC
The following table shows the components that are installed with the QC_Protocols configuration.
Components Installed with QC_Protocols
Component
Items Installed
Protocols
DNA Initial QC 5.1.2
Library Validation QC 5.1.2
RNA Initial QC 5.1.2
Master Steps
Add Multiple Reagents
Adenylate ends & Ligate Adapters (TruSeq DNA) 5.1
Pool Samples
Aggregate QC (DNA) 5.1.2
Aggregate QC (Library Validation) 5.1.2
Aggregate QC (RNA) 5.1.2
Bioanalyzer QC (DNA) 5.1.2
Bioanalyzer QC (RNA) 5.1.2
CaliperGX QC 5.1.2
NanoDrop QC (DNA) 5.1.2
NanoDrop QC (RNA) 5.1.2
PicoGreen QC (DNA) 5.1.2
PicoGreen QC (RNA) 5.1.2
Qubit QC (DNA) 5.1.2
Qubit QC (RNA) 5.1.2
Tapestation QC (DNA) 5.1.2
Tapestation QC (RNA) 5.1.2
qPCR QC 5.1.2
Container types
384 well plate
96 well plate
BioAnalyzer DNA 1000 Chip
BioAnalyzer DNA High Sensitivity Chip
BioAnalyzer RNA Nano Chip
BioAnalyzer RNA Pico Chip
Tube
Label groups/reagent categories
TruSeq Custom Amplicon Adapters (A7-A5)
TruSeq DNA LT Adapters (AD series)
TruSeq Stranded mRNA LT Adapters (AR series)
TruSeq HT Adapters (D7-D5)
Nextera Enrichment & Rapid Capture Adapters (N7-E5)
Nextera DNA Adapters (N7-N5, S7-S5)
NoIndex
TruSeq Small RNA Adapters (RPI series)
Reagent kits
TruSeq DNA PCR-Free Kit - Adapter Plate Box
TruSeq DNA PCR-Free Kit - Core Reagents Box
TruSeq DNA PCR-Free Kit - SP Beads Box
Controls
Endogenous Positive Control
Exogenous Positive Control
No Amplification Control
No Reverse Transcriptase Control
No Template Control
PhiX v3
Sample custom fields
Application
Pooling
Progress
Reference Genome
Sample Buffer
Sample Conc.
Units
Volume (µl)
Read Length
Sample Type
Sequencing Coverage
Sequencing Method
Container custom fields
Container Name
LIMS ID (Container)
Well
Project custom fields
Project Name
Derived sample custom fields
A260/280 ratio
Conc. Units
Concentration
Sample Name
Waiting
Workflow
Measurement custom fields
28s/18s ratio
A260
A260/230 ratio
A260/280 ratio
BA Sample Name
Conc. Units
Concentration
MW Units
Molarity Units
Number of Peaks found
Number of Regions found
Peak 1% Integrated Area
Peak 1 Conc.
Peak 1 MW
Peak 1 Molarity
Peak 1 Size - bp
Peak 2% Integrated Area
Peak 2 Conc.
Peak 2 MW
Peak 2 Molarity
Peak 2 Size - bp
Peak 3% Integrated Area
Peak 3 Conc.
Peak 3 MW
Peak 3 Molarity
Peak 3 Size - bp
Peak 4% Integrated Area
Peak 4 Conc.
Peak 4 MW
Peak 4 Molarity
Peak 4 Size - bp
Peak 5% Integrated Area
Peak 5 Conc.
Peak 5 MW
Peak 5 Molarity
Peak 5 Size - bp
RIN
RNA Area
Region 1% of Total
Region 1 Average Size - bp
Region 1 Conc.
Region 1 Molarity
Region 2% of Total
Region 2 Average Size - bp
Region 2 Conc.
Region 2 Molarity
Region 3% of Total
Region 3 Average Size - bp
Region 3 Conc.
Region 3 Molarity
Region 4% of Total
Region 4 Average Size - bp
Region 4 Conc.
Region 4 Molarity
Region 5% of Total
Region 5 Average Size - bp
Region 5 Conc.
Region 5 Molarity
Sample Comment
Size (bp)
The following table shows the AmpliSeq_for_Illumina_Protocols workflow configuration package files and the workflows that are installed with IPP Integration v2.10.
AmpliSeq_for_Illumina_Protocols Configuration Package Files and Workflows
Configuration Package Files
Workflows
AmpliSeq-Illumina-BRCA-Panel-v1.1
Equalizer Workflow AmpliSeq for Illumina BRCA Panel v1.1
Library Prep AmpliSeq for Illumina BRCA Panel v1.1
Standard Workflow AmpliSeq for Illumina BRCA Panel v1.1
AmpliSeq-Illumina-Cancer-HotSpot-Panel-V2-v1.1
Library Prep AmpliSeq for Illumina Cancer HotSpot Panel V2 v1.1
Standard Workflow AmpliSeq for Illumina Cancer HotSpot Panel V2 v1.1
AmpliSeq-Illumina-Childhood-Cancer-Panel-v1.1
DNA Library Prep AmpliSeq for Illumina Childhood Cancer Panel v1.1
Equalizer Workflow AmpliSeq for Illumina Childhood Cancer Panel v1.1
RNA Library Prep AmpliSeq for Illumina Childhood Cancer Panel v1.1
Standard Workflow AmpliSeq for Illumina Childhood Cancer Panel v1.1
AmpliSeq-Illumina-Comprehensive-Cancer-Panel-v1.1
Equalizer Workflow AmpliSeq for Illumina Comprehensive Cancer Panel v1.1
Library Prep AmpliSeq for Illumina Comprehensive Cancer Panel v1.1
Standard Workflow AmpliSeq for Illumina Comprehensive Cancer Panel v1.1
AmpliSeq-Illumina-Comprehensive-Panel-v3-v1.1
DNA Library Prep AmpliSeq for Illumina Comprehensive Panel v3 v1.1
Equalizer Workflow AmpliSeq for Illumina Comprehensive Panel v3 v1.1
RNA Library Prep AmpliSeq for Illumina Comprehensive Panel v3 v1.1
Standard Workflow AmpliSeq for Illumina Comprehensive Panel v3 v1.1
AmpliSeq-Illumina-Custom-DNA-Panel-v1.1
Equalizer Workflow AmpliSeq for Illumina Custom DNA Panel v1.1
Library Prep AmpliSeq for Illumina Custom DNA Panel v1.1
Standard Workflow AmpliSeq for Illumina Custom DNA Panel v1.1
AmpliSeq-Illumina-Focus-Panel-v1.1
DNA Library Prep AmpliSeq for Illumina Focus Panel v1.1
Equalizer AmpliSeq for Illumina Focus Panel v1.1
RNA Library Prep AmpliSeq for Focus Panel v1.1
Standard Workflow AmpliSeq for Illumina Focus Panel v1.1
AmpliSeq-Illumina-Immune-Repertoire-Panel-v1.1
Equalizer AmpliSeq for Illumina Immune Repertoire Panel v1.1
RNA Library Prep AmpliSeq for Illumina Immune Repertoire Panel v1.1
Standard Workflow AmpliSeq for Illumina Immune Response Panel v1.1
AmpliSeq-Illumina-Immune-Response-Panel-v1.1
Equalizer Workflow AmpliSeq for Illumina Immune Response Panel v1.1
RNA Library Prep AmpliSeq for Illumina Immune Response Panel v1.1
Standard Workflow AmpliSeq for Illumina Immune Response Panel v1.1
AmpliSeq-Illumina-Myeloid-Panel-v1.1
DNA Library Prep AmpliSeq for Illumina Myeloid Panel v1.1
Equalizer Workflow AmpliSeq for Illumina Myeloid Panel v1.1
RNA Library Prep AmpliSeq for Illumina Myeloid Panel v1.1
Standard Workflow AmpliSeq for Illumina Myeloid Panel v1.1
AmpliSeq-Illumina-TCR-beta-SR-Panel-v1.1
DNA Library Prep AmpliSeq for Illumina TCR beta-SR Panel v1.1
RNA Library Prep AmpliSeq for Illumina TCR beta-SR Panel v1.1
AmpliSeq-Illumina-Transcriptome-Human-Gene-Expression-Panel-v1.1
Equalizer Workflow AmpliSeq for Illumina Transcriptome Human Gene Expression Panel v1.1
RNA Library Prep AmpliSeq for Illumina Transcriptome Human Gene Expression Panel v1.1
Standard Workflow AmpliSeq for Illumina Transcriptome Human Gene Expression Panel v1.1
The following table shows additional workflow configuration package files and workflows that are installed with IPP Integration v2.10.
Additional Configuration Package Files and Workflows
Configuration Package File
Workflows
Covid
CDC COVID-19 RT PCR v1.1
Qiagen EZ1 Advanced XL CDC COVID-19 RT-PCR v1.1
Qiagen QIAamp DSP Viral RNA Mini Kit CDC COVID-19 RT-PCR v1.1
Roche MagNA Pure 96 CDC COVID-19 RT-PCR v1.1
Roche MagNA Pure Compact CDC COVID-19 RT-PCR v1.1
Roche MagNA Pure LC CDC COVID-19 RT-PCR v1.1
Sort Specimens to Extraction CDC COVID-19 RT-PCR v1.1
bioMerieux EMAG Instrument CDC COVID-19 RT-PCR v1.1
bioMerieux NucliSENS easyMAG Instrument CDC COVID-19 RT-PCR v1.1
Illumina COVIDSeq Workflow v1.6
Respiratory Virus Panel v1.0
DNA_Protocols
Illumina DNA PCR-Free Library Prep Manual v1.0.3
Illumina DNA Prep (M) Tagmentation v1.0.10
Nextera Mate Pair v1.0.11
Nextera XT DNA v2.0.11
TruSeq DNA PCR-Free v2.0.11
TruSeq Nano DNA v1.0.11
IGA
Illumina Genomics Architecture - Library Prep Automated v2.1
Sample Selection (IGA v2.1)
Sample Plating (IGA v2.1)
Blood Extraction (IGA v2.1)
Saliva Extraction (IGA v2.1)
Illumina DNA PCR-Free (IGA v2.1)
Illumina DNA with Enrichment (IGA v2.1)
Sequencing Platform Selection (IGA v2.1)
Illumina Genomics Architecture - NovaSeq Sequencing v2.1
Run Format (IGA v2.1)
NovaSeq Standard (IGA v2.1)
NovaSeq Xp (IGA v2.1)
AUTOMATED - NovaSeq Run (IGA v2.1)
ℹ IGA workflows require additional installation steps. For assistance, contact Illumina Technical Support. Note also that Illumina Genomics Architecture - NovaSeq Sequencing v2.1 workflow is not the same workflow as NovaSeq 6000 v3.8.
Infinium_Array
GenomeStudio v1.0
ILASS Infinium Batch DNA v1.1
ILASS Infinium Genotyping Assay v1.1
ILASS Infinium Genotyping with PGx v1.1
Infinium HD Methylation Assay Manual v1.2
Infinium HTS Assay Manual v1.2
Infinium LCG Assay Manual v1.2
Infinium XT Assay Manual v1.2
Methylation_Protocols
TruSeq ChIP-Seq v1.0.11
TruSeq Methyl Capture EPIC v2.0.12
RNA_Protocols
Illumina Stranded mRNA Prep Ligation v1.1
Illumina Stranded Total RNA Prep Ligation with Ribo-Zero Plus v1.1
TruSeq Small RNA v1.0.11
TruSeq Stranded mRNA v2.1.11
Targeted_Amplicon_Protocols
TruSeq Custom Amplicon v1.0.11
Targeted_Enrichment_Protocols
Illumina DNA Prep with Enrichment (S) Tagmentation v1.2
Illumina RNA Prep with Enrichment (L) Tagmentation v1.1
Nextera Rapid Capture Custom Enrichment v2.0.12
TruSeq DNA Exome v2.0.11
TruSeq RNA
TruSeq RNA Access v2.0.11
TruSeq RNA Exome v1.0.7
TruSight_Oncology_Protocols
TruSight Oncology 500 ctDNA v1.1
TruSight Oncology 500 HT v1.1
TruSight Oncology 500 v1.1
TruSight Tumor 170 v2.0.11
The BaseSpaceLIMS-Ilumina-Preset-Protocols RPM package provides instructions for locating and running the /opt/gls/clarity/config/illumina-preset-protocols-installer.sh bash script that launches the IPP configuration installer.
The RPM package automatically installs template files that are used with Template File Generator (TFG) and Sample Placement Helper. These template files are accessible after the applicable configuration is installed. The files are installed at the following locations:
/opt/gls/clarity/extensions/conf/driverfiletemplates
/opt/gls/clarity/extensions/conf/infinium/placementpatterns
/opt/gls/clarity/extensions/conf/covid
The RPM package automatically installs the CLPA automation that is used for the CLPA integration. This automation is installed at the following location:
After you have selected a workflow, the installer validates the import of that workflow and provides Warning/Error details in STDOUT. This process allows you to proceed with the import or cancel it.
The IPP Integration v2.10 installer uses parameters and operations to assist with the configuration installation.
You must run the installer as the glsjboss user. The following table describes the parameters that are available for use with the installer.
IPP Installer Parameters
Parameter
Description
Example Command
--operation, -o
Specifies the run mode for the installer. Supported modes are as follows:
help
list
install
--headless, -h
Run the installer in headless mode. Used for the install operation only.
--log, -l
Sets the log file name and destination. The default value is ipp-installer.log.
Run the installer with the list operation for a list of all configuration installation options in a two-column format. The first column shows the IPP workflow identifier (or Id) that the installer uses when running the install operation. The second column shows the name of the configuration associated with the identifier. As the glsjboss user, run the following command to view the list of IPP workflows:
Use the install operation to install the workflow configuration package file. Specify both the name of the parent item and the name of the workflow configuration package file, separated by a period, as follows:
For example, to install the qc-protocols workflow configuration package file, run the following command:
All Operation
The all operation is a special option that is available at the top level and parent item level.
At the top level, use the operation to install every configuration package from IPP Integration v2.10. To use the all operation at this level, run the following command:
At the parent item level, use the operation to install the workflows associated with a specific parent item. For example, run the following command to install the workflows associated with the AmpliSeq_for_Illumina_Protocols parent item:
There is no all option available for QC_Protocols and Targeted_Amplicon_Protocols as they each contain only one workflow configuration package file. Including all in the install command for these items generates an error.
Headless Mode
Headless mode allows the install operation to complete without prompting for input. Use this mode if you want to automate the installation. When running the installation in headless mode, the process still runs through the validation phase before importing the configuration. However, if a conflict is found, the configuration causing the conflict is skipped automatically.
The IPP installer validates the configuration import. If conflicts are found during the validation, a warning message displays. This message provides the following options:
f/Import anyway
s/Skip this workflow
a/Abort
The f/Import anyway option runs the import command of the config slicer tool. It also allows the tool to handle conflicts. This option does not run the config slicer tool in importAndOverwrite mode.
The s/Skip this workflow option does not import workflows where conflicts were found, but continues to import the remaining workflow package files (if any).
The a/Abort option aborts all import operations. The conflicts found are captured in the log files and can be reviewed.
The IPP installer tracks the installation process in the installhistory table and generates log files that can be reviewed after the installation is complete.