Manual Upgrade

NextSeq 1000/2000 Integration v2.5.0 supports future ICA services and includes new custom fields and modifications to existing automations. The integration package requires the NextSeq 1000/2000 Sequencing v2.4 workflow. You can install this workflow through the Illumina Preset Protocols (IPP). If you do not want to upgrade through the IPP, you can manually upgrade the workflow configuration.

hashtag
Modify the Existing NextSeq 1000/2000 Workflow

To modify the existing NextSeq 1000/2000 Sequencing v2.3 workflow, you must make the following changes:

  • Add the Measurement global field.

  • Modify the Master Step field.

  • Modify the Demultiplexing (NextSeq 1000/2000 Sequencing v2.3) master and protocol steps.

  • Update the Validate Reagent Cartridge Barcode step automation.

chevron-rightAdd Index Sequence Global Fieldhashtag
  1. From Configuration, select the Custom Fields tab.

  2. In Global Fields, create a new Measurement field with the following properties:

chevron-rightModify Master Step Fieldhashtag
  1. From Configuration, select the Custom Fields tab.

  2. Select the Master Step Fields tab.

chevron-rightModify Demultiplexing (NextSeq 1000/2000 Sequencing v2.3) Master Stephashtag
  1. From Configuration, select the Lab Work tab.

  2. Navigate to the Steps pane and select Demultiplexing (NextSeq 1000/2000 Sequencing v2.3).

chevron-rightModify Demultiplexing (NextSeq 1000/2000 Sequencing v2.3) Protocol Stephashtag
  1. From Configuration, select the Lab Work tab.

  2. Navigate to the Steps pane and select Demultiplexing (NextSeq 1000/2000 Sequencing v2.3).

chevron-rightUpdate Validate Reagent Cartridge Barcode Step Automationhashtag
  1. From Configuration, select the Automation tab, and then select the Step Automation tab.

  2. Select Validate Reagent Cartridge Barcode.

hashtag
Modify NextSeq 1000/2000 Sequencing v2.5.0 Integration Package Configuration

When the /opt/gls/clarity/config/configure_nextseq1k2k.sh script is running as part of the installation and configuration process, do not use the default names for the automated sequencing and demultiplexing steps. Instead, use the names in the modified workflow.

circle-info

This integration moves the integration service and integration-related properties from the configuration files to the database.

To make modifications after configuration, use the omxProps-ConfigTool utility to update the following properties:

Field Name — Index Sequencing

  • Field Type — Text

  • Required Field — No

  • Read Only — Yes

  • Select Save.

  • Select Demultiplexing (NextSeq 1000/2000 Sequencing v2.3) to display the applicable fields.

  • Select Criteria 1 - Source Data Field.

  • In the Additional Options pane, remove # of > = Q30 Bases (PF) from the Dropdown Items list.

  • Select Criteria 2 - Source Data Field.

  • In the Additional Options pane, remove # of > = Q30 Bases (PF) from the Dropdown Items list.

  • Select Save.

  • Select the Master Step tab, and then select Record Details.

  • In the Step File Placeholders pane, rename Demultiplex Stats to Run Metrics.

  • Select Save.

  • Select the Step tab, and then select Record Details.

  • Add the Analysis Status field as follows.

    1. In the Step Data pane, select + to open the Master Step Fields pane.

    2. Select Analysis Status, and then select the check mark to add the field to the Master Step Fields list.

  • Update the Sample Table as follows.

    1. In the Sample Table pane, navigate to the Table Columns pane and remove the following fields:

      • [Measurement] # of > = Q30 Bases (PF)

      • [Measurement] Mean Quality Score (PF)

    2. Select + to open the Fields pane.

    3. Select Measurement to open the field drop-down list.

    4. Select Index Sequence.

    5. Reorder the table columns as follows.

      • [Derived Sample] Sample Name

      • [Measurement] Index Sequence

    6. Select Save.

  • In Automation Details, replace the content of the Command Line field with the following command:

  • Select Save.

  • Property

    Description

    integration.nextseq1k2k.v2.automatedStepNames

    Comma-separated step names for the automated sequencing step

    integration.nextseq1k2k.v2.analysisStepNames

    Comma-separated step names for the demultiplexing step

    bash -l -c "/opt/gls/clarity/bin/java -jar /opt/gls/clarity/extensions/ngs-common/v5/EPP/ngs-extensions.jar -i {stepURI:v2} -u {username} -p {password} script:evaluateDynamicExpression -exp 'if (!output.container.name.matches( ::[A-Z]{2}[0-9]{7}-[A-Z0-9]{4}:: ) ) {fail ( ::Invalid Reagent Cartridge Barcode. Please verify and try again.:: ) }' -log {compoundOutputFileLuid1}"
    [Container] Well
  • [Measurement] # Reads

  • [Measurement] # Perfect Index Reads

  • [Measurement] # One Mismatch Index Reads

  • [Container] Container Name

  • [Project] Project Name