Unexpected Variant Calls
Last updated
Was this helpful?
If you observe unexpected variant calls that you suspect are artifacts, check the following:
Verify that you provided a systematic noise file as input, especially in tumor-only mode.
Confirm that the samples used to build the noise baseline match your tumor samples in:
Sample type (FFPE, fresh frozen, blood)
Sample quality (DIN score range)
Library prep kit
Use at least 30 samples for the baseline. See Stage 1: Generate a Systematic Noise File for best practices.
Check the Supplementary (chimeric) alignments QC metric. If the value is elevated above 1–2%, or above your typical range, consider running the sample in low-quality FFPE mode.
If none of the above resolves the issue, contact Illumina Technical Support at [email protected].
Last updated
Was this helpful?
Was this helpful?
