Optical Target Test
Before processing images for the Illumina Spatial Transcriptome Assay, the Optical Target Test should be performed using the Illumina Bright Field (BF) Optical Target. This test is intended to evaluate whether the image quality produced by a microscope is suitable for use with the Illumina Spatial Transcriptome Assay. The images captured during this test are used as input for the Illumina Spatial Image Tool (ISIT) during the Assess Microscope workflow.
The Optical Target Test provides the following key metrics:
Estimated pixel size (microns) of the camera
Full Width Half Maximum (FWHM) for assessing focus quality
Shading uniformity observed in the FOV
An optical distortion heatmap
The Illumina BF Optical Target is a 75 mm x 25 mm x 1 mm slide featuring patterned elements in the center. This test must be conducted using the 20X objective lens and the specific camera intended to be used with the Illumina Spatial Transcriptome Assay. The microscope assessment includes evaluation of focus quality, detection of stage tilt, analysis of optical distortion, and assessment of shading uniformity. These metrics are valuable for diagnosing and troubleshooting image quality issues.