DRAGEN TSO 500 Analysis Software has optional and required fields that are required in addition to general sample sheet requirements. Follow the steps below to create a valid samplesheet.
The following sample sheet requirements describe required and optional fields for DRAGEN TSO 500 Analysis Software. Depending on the deployment (standalone DRAGEN server, ICA with auto-launch, ICA with manual launch), certain sections and required values can deviate from the standard requirements. These deviations are noted in the information below.
The analysis fails if the sample sheet requirements are not met.
Use the following steps to create a valid sample sheet.
Download the sample sheet v2 template that matches the instrument & assay run.
In the BCL Convert Settings section, enter the following required parameters:
In the BCL Convert Data section, enter the following parameters for each sample.
In the TSO 500 Data section, enter the following parameters:
To ensure a successful analysis, follow these guidelines:
Avoid any blank lines at the end of the sample sheet; these can cause the analysis to fail.
When running local analysis using the command line save the sample sheet in the sequencing run folder with the default name SampleSheet.csv
, or choose a different name and specify the path in the command-line options.