Troubleshooting

hashtag
General Troubleshooting on Standalone DRAGEN Server

Failure Type
Actions

Software

- Open the log file ./<AnalysisFolder>/Logs_Intermediates/pipeline_trace.txt. This log file displays each pipeline step run by the Nextflow workflow manager software. If a step fails, it is marked as FAILED. Each step generates log files that are stored in step-specific subfolders in the Logs_Intermediates folder. Review the log files in the relevant Logs_Intermediates folder for the step to identify potential sources of error. - Open the errors folder ./<AnalysisFolder>/errors. The workflow creates an error file, error_<NameOfFailedStep>.json, for each step that failed during analysis. For steps that fail per sample, there is a separately labeled file for each sample that failed each step error_<NameOfFailedStep>_<SampleIDIfRelevant>.json. These files contain the command and stdout and stderr from the step.

Samples

hashtag
Sample Sheet Validation Failures

In DRAGEN TruSight Oncology 500 ctDNA Analysis Software, the analysis fails if a sample sheet is invalid. If an invalid sample sheet in suspected, log files can help troubleshoot a failed analysis. Use the following steps to find the log file for the sample sheet:

  1. Navigate to the following location /<analysis_output>/Logs_Intermediates/SamplesheetValidation.

  2. Open the SamplesheetValidation-.log file

  3. Find a line with the following: SampleSheetValidationTask:NA:1 exited with return code 1 which has not been declared as a valid return code.

hashtag
General troubleshooting for a failed sample sheet:

Failure Type
Action

hashtag
Valid indexes for assay and index combinations:

Assay
Index Set ID

hashtag
Troubleshooting BCL issues:

Failure Type
Action

hashtag
Troubleshooting FASTQs issues:

Failure Type
Action

hashtag
Other Troubleshooting Tips

Failure Type
Action

hashtag
Troubleshooting on ICA

In addition to TSO 500 managed sample sheet validations, ICA managed TSO 500 errors include the following:

Error
Description
  • Search for errors in the sample sheet validation log and compare with the guidelines and warnings in and the following tables.

  • Open the combined metrics output results file ./<AnalysisFolder>/Results/<PairId>/MetricsOutput.tsv. If a sample fails an analysis step, the Pair ID that contains the sample shows the failure under FAILED_STEPS in the Analysis Status section, and COMPLETED_ALL_STEPS shows as False. If available, review the individual log files for the failed steps under ./<AnalysisFolder>/Logs_Intermediates to identify potential sources of error.

    Sample Sheet not found

    Verify that SampleSheet.csv is present at the top level of the run folder with the name "SampleSheet.csv". If the sample sheet is in a different location, supply the sample sheet using the --sampleSheet option

    Indexes are not valid for the sequencer and/or assay

    See Valid indexes for assay and instrument combinations for correct indexes for the sequencer and assay.

    Sample Sheet is not in v2 format

    Verify that the format of the sample sheet is v2. v1 sample sheet is not compatible with DRAGEN TruSight Oncology 500 ctDNA Analysis Software.

    Analysis does not run

    Verify the analysis starts from the run folder, and BCLs or FASTQs are in the correct locations as outlined in Starting From BCL Files and Starting From FASTQ Files respectively.

    TSO 500 ctDNA

    • UP1-UP16

    TSO 500 ctDNA v2

    • UDP0001–UDP0192

    Lane Column without Values

    Ensure that the column is completed. If lane is not applicable to the run, delete the column.

    Format of v2 sample sheet is incorrect

    Verify that the sections and fields are present in the sample sheet and follow the individual rules in

    Unique sample IDs

    Verify that the Sample_IDs are unique in the sample sheet.

    Incorrect folder structure

    Verify that the FASTQ files are in the correct structure. Refer to for more information.

    Invalid FASTQ input files

    If the FASTQs are invalid, start TSO 500 ctDNA analysis from BCL files.

    The output file directory contains information from previous analyses

    If this issue is seen: specify a new target output folder and repeat analysis To prevent this issue: specify an empty directory before starting analysis

    Single exon (single probe) genes are still reported in the CNV VCF file, but not the CNV TSV file

    No action needed; software is working as expected.

    Currently all single probe genes are not emitted to the Copy Number Variants section of our CombinedVariantOutput.tsv. However, you can still find these events in the cnv.vcf.gz.

    Due to the single probe nature, accurate CNV calling has not been validated and as such they are emitted as REF

    Testing of cell lines, contrived samples and commercial controls does not return expected results

    Review recommendations for using these samples types .

    Failure type: ValueError: Could not find pipeline ID for app BCLConvert in sample sheet SampleSheet.csv

    Action: Ensure StartsFromFastq field is in the [TSO500L_Settings] section, and it is not present in the [BCLConvert_Settings] Section. Refer to for more information.

    Sample Sheet Requirements
    Sample Sheet Requirements
    Starting From FASTQ Files
    here
    Sample Sheet Requirements