Analysis Launch on Standalone DRAGEN Server
Start the DRAGEN TruSight Oncology 500 ctDNA Analysis Software with the DRAGEN_TSO500_CTDNA-2.6.0.sh
Bash script. The script is installed in the /usr/local/bin directory
. The Bash script is executed on the command line and runs the software with Docker (or Apptainer if specified).
For arguments, refer to Command-Line Options. You can start from BCL files or from the FASTQ folder produced by BCL Convert. The following requirements apply for both methods:
Path to the sequencing run or FASTQ folder. Copy the run or FASTQ folder to the DRAGEN server into the staging folder with the following recommended organization:
/staging/runs/{RunID}
. You can copy the run folder onto the DRAGEN server using Linux commands such asrsync
. The sample sheet within the run folder is used unless otherwise specified through the command line.Run folder must be intact. Refer to Starting from BCL Files for input requirements.
If the analysis output folder path is different from the default, provide the analysis output folder path. Refer to Command-Line Options.
Before running the analysis, confirm that the output directory for the software to write to is empty and does not include results of previous analyses.
Storage Requirements
For optimal performance, run analysis on data stored locally on the DRAGEN server. Analysis of data stored on NAS can take longer and performance can be less reliable.
The DRAGEN server provides an NVMe SSD in the /staging directory to use as the software output directory. Network-attached storage is required for long-term storage.
When running the DRAGEN TruSight Oncology 500 ctDNA Analysis Software, use the default settings or set the -analysisFolder command line option to a directory in /staging to make sure the DRAGEN server processes read and write data on the NVMe SSD.
Before beginning analysis, develop a strategy to copy data from the DRAGEN server to a network‑attached storage. Delete output data on the DRAGEN server as soon as possible.
The following are the run and analysis output sizes for each sequencing system per 101 bp:
NovaSeq 6000/6000Dx (RUO) SP Flow Cell
85-100
250-374
300
NovaSeq 6000/6000Dx (RUO) S1 Flow Cell
164-200
360-665
800
NovaSeq 6000/6000Dx (RUO) S2 Flow Cell
290-460
890-1600
1500
NovaSeq 6000/6000Dx (RUO) S4 Flow Cell
800-1200
2700-4100
3000
NovaSeq X 1.5B
213
352
800
NovaSeq X 10B
1100
1800
3000
When launching the analysis, the software checks that the minimum disk space required is available. If the minimum disk space is not available, the software shows an error message and prevents analysis from starting. If disk space is exhausted during a run, the run shows an error and stops analyzing.
Moving or modifying files during an analysis may cause the analysis to fail or provide incorrect results.