Troubleshooting
Failure Type | Actions |
---|---|
Software | - Open the log file |
Samples | Open the combined metrics output results file |
Multinode Gather | If the following error appears, check if the sample or pair ID was included multiple times during separate node analysis runs, before being gathered together. If the error exists, rerun one of the analyses without the duplicate and reattempt gathering. |
Sample Sheet Validation Failures
In DRAGEN TruSight Oncology 500 Analysis Software, the analysis fails if a sample sheet is invalid. If an invalid sample sheet in suspected, log files can help troubleshoot a failed analysis. Use the following steps to find the log file for the sample sheet:
Navigate to the following location
/<analysis_output>/Logs_Intermediates/SamplesheetValidation.
Open the
SamplesheetValidation-.log
fileFind a line with the following:
SampleSheetValidationTask:NA:1 exited with return code 1 which has not been declared as a valid return code.
Search for errors in the sample sheet validation log and compare with the guidelines and warnings in [Sample Sheet Requirements](add link) and the following tables.
General troubleshooting for a failed sample sheet
Failure Type | Action |
---|---|
Sample Sheet not found | Verify that |
Indexes are not valid for the sequencer and/or assay | See Valid indexes for assay and instrument combinations for correct indexes for the sequencer and assay. |
Pair_ID is not unique | Pair_ID column is required in the TSO500S_Data section of the sample sheet, which pairs at most one RNA and one DNA sample together for analysis. If the sample does not have a pair, use a unique pair ID for single samples. |
Sample Sheet Not in v2 Format | Verify that the format of the sample sheet is in v2 format. v1 sample sheet is not compatible with DRAGEN TruSight Oncology 500 Analysis Software. |
Analysis does not run | Verify the analysis starts from the run folder, and BCLs or FASTQs are in the correct locations as outlined in Starting From BCL Files and Starting From FASTQ Files respectively. |
Valid indexes for assay and instrument combinations
Assay | NextSeq 550 | NextSeq 550 |
---|---|---|
TSO 500 | - UP1-UP16 - CP1-CP16 (DNA Only) | - UP1–UP16 |
TSO 500 HT | UDP0001–UDP0192 | UDP0001–UDP0192 |
Troubleshooting BCL issues
Failure Type | Action |
---|---|
Lane Column without Values | Ensure that the column is completed. If lane is not applicable to the run, delete the column. |
Format of v2 sample sheet is incorrect | Verify that the following sections and fields are present in the sample sheet and follow the individual rules in [Sample Sheet Requirements](Add Link to SS requirements). [BCLConvert_Settings] - SoftwareVersion - AdapterRead1 - AdapterRead2 - AdapterBehavior - MinimumTrimmedReadLength - MaskShortReads [BCLConvert_Data] - Sample_ID - index - index2 [TS0500S_Data] - Sample_ID - Index_ID - Sample_Type - Pair_ID |
HRD analysis is missing | Make sure that HRD is in the correct column in the sample sheet. Refer to Sample Sheet Requirements for more information. |
Troubleshooting FASTQ issues
Failure Type | Action |
---|---|
Sample_ID and/or Sample_Type is not present | Verify that the sample sheet has columns and values for Sample_ID and Sample_Type. |
Unique sample IDs | Verify that the Sample_IDs are unique in the sample sheet. |
Format of v2 Sample Sheet is incorrect | Verify that the following sections and fields are present in the sample sheet and follow the individual rules in Sample Sheet Requirements. [TS0500S_Data] - Sample_ID - Index_ID - Sample_Type - Pair_ID Make sure that HRD, if needed, is in the correct column in the sample sheet. Refer to Sample Sheet Requirements for more information. |
Incorrect folder structure | Verify that the FASTQ files are in the correct structure. Refer to Starting From FASTQ Files for more information. |
Invalid FASTQ input files | If the FASTQs are invalid, start TSO 500 analysis from BCL files. |
HRD analysis missing | Make sure that HRD is in the correct column in the sample sheet. |