Release Notes
Last Updated: November 2024
Release Date: September 2023
Document Version: 2
These release notes describe the key changes to software components for the Clarity LIMS NextSeq 1000/2000 Integration Package v2.4.0.
Compatibility
Refer to Compatibility under Instruments & Integrations.
New Features
Updates third-party dependency libraries.
Defect Repairs
The NextSeq 1000/2000 sample sheet now reflects the submitted sample name instead of the library name.
The scope for the JWT token in the integration is restricted.
Fixed SIS core service not being properly registered in systemctl.
Known Issues
Sequencing run error triggered by disk space full on instrument causes RunParameters.xml to be empty and the run event cannot be processed on Clarity LIMS. This error is unlikely to occur as pre-run checks on the instrument check for sufficient disk space before run is started.
Sample sheet and planned run generation will fail if any of the samples in the pools has been assigned QC flag in prior steps before entering the Load to Reagent Cartridge step.
Limitations
For the demultiplexing information in the Sample Details table to display correctly, the following conditions must be met:
The DRAGEN version must be earlier than v3.9.3.
The samples in the run must not repeat.
The run must be for the single lane flow cell (P1 or P2).
Revision History
Last updated